| Assessment Year | 1993 |
|---|---|
| Principal Product | DISPOSABLE PLASTIC TUBES |
| NAICS |
n/a
|
| SIC | 3841: Surgical and Medical Instruments and Apparatus |
| Sales Range | $10 M to 50M |
| # of Employees | 715 |
| Plant Area |
n/a
|
| Annual Production | 207,000,000
pieces
|
| Production Hrs. Annual | 7,680
hrs
|
| Location (State) | Missouri |
Facility Energy Cost and Usage
| Source | Yearly Cost | Usage | Unit Cost | |
|---|---|---|---|---|
| Electricity Usage | $583,424 | 14,049,531 |
kWh
|
$0.042 |
| Natural Gas | $8,756 | 2,183 |
MMBtu
|
$4.01 |
| TOTAL YEARLY COSTS | - | |||
| RECOMMENDED SAVINGS* | $49,794 |
See individual recommendations below
*Non-energy impacts included in savings.
|
||
| IMPLEMENTED SAVINGS* | - | |||
Recommendations
| # | Description | Yearly Savings* | Imp'd Cost* | Status | Electricity Usage Savings | Natural Gas Savings | Administrative Changes Savings |
|---|---|---|---|---|---|---|---|
| 01 | 2.7142: Utilize higher efficiency lamps and/or ballasts | $7,521 | - |
$9,089
218,933
kWh
|
- | $(-1,568.0) | |
| 02 | 2.7135: Install occupancy sensors | $18,864 | $8,935 |
$18,864
454,279
kWh
|
- | - | |
| 03 | 2.7142: Utilize higher efficiency lamps and/or ballasts | $6,058 | - | - | - | $6,058 | |
| 04 | 2.7111: Reduce illumination to minimum necessary levels | $2,726 | $400 |
$2,726
65,651
kWh
|
- | - | |
| 05 | 2.7141: Lower light fixtures in high ceiling areas | $7,132 | $6,200 |
$7,132
171,747
kWh
|
- | - | |
| 06 | 2.2511: Insulate bare equipment | $4,454 | $3,772 |
$4,454
107,268
kWh
|
- | - | |
| 07 | 2.4221: Install compressor air intakes in coolest locations | $2,762 | $990 |
$2,762
66,530
kWh
|
- | - | |
| 08 | 2.7261: Install timers and/or thermostats | $277 | $1,000 |
$189
4,396
kWh
|
$88
-
MMBtu
|
- | |
| TOTAL RECOMMENDED | $49,794 | $21,297 |
$45,216
1,088,804
kWh
|
$88
-
MMBtu
|
$4,490 | ||
| TOTAL IMPLEMENTED | - | - | - | - | - | ||
| Yearly Savings* | Imp'd Cost* | Electricity Usage Savings | Natural Gas Savings | Administrative Changes Savings | |||
*Savings and costs are best estimates based on the center's site visit, associated engineering analysis, and typical capital and labor costs