Assessment Year | 2017 |
---|---|
Principal Product | Medical Devices |
NAICS | 339112: Surgical and Medical Instrument Manufacturing |
SIC | 3841: Surgical and Medical Instruments and Apparatus |
Sales Range | $50 M to 100M |
# of Employees | 250 |
Plant Area |
137,945
ft2
|
Annual Production | 200,000
units
|
Production Hrs. Annual | 2,730
hrs
|
Location (State) | Utah |
Facility Energy Cost and Usage
Source | Yearly Cost | Usage | Unit Cost | |
---|---|---|---|---|
Electricity Usage | $94,833 | 2,581,800 |
kWh
|
$0.037 |
Electricity Demand | $69,841 | 5,530 |
kW-mo/yr
|
$12.63 |
Electricity Fees | $37,920 | - |
|
- |
Natural Gas | $77,639 | 12,126 |
MMBtu
|
$6.40 |
TOTAL YEARLY COSTS | $280,232 | |||
RECOMMENDED SAVINGS* | $24,096 |
See individual recommendations below
*Non-energy impacts included in savings.
|
||
IMPLEMENTED SAVINGS* | $7,381 |
Recommendations
# | Description | Yearly Savings* | Imp'd Cost* | Status | Electricity Usage Savings | Electricity Demand Savings | Electricity Fees Savings |
---|---|---|---|---|---|---|---|
01 | 2.8121: Apply for tax-free status for energy purchases | $13,592 | $500 | - | - | $13,592 | |
02 | 2.2211: Use optimum temperature | $7,381 | - |
$3,992
108,798
kWh
|
$2,566
-
kW-mo/yr
|
$823 | |
03 | 2.4231: Reduce the pressure of compressed air to the minimum required | $2,762 | - |
$1,703
46,371
kWh
|
$802
-
kW-mo/yr
|
$257 | |
04 | 2.7144: Install spectral reflectors / delamp | $361 | $113 |
$226
6,161
kWh
|
$28
-
kW-mo/yr
|
$107 | |
TOTAL RECOMMENDED | $24,096 | $613 |
$5,922
161,330
kWh
|
$3,396
-
kW-mo/yr
|
$14,778 | ||
TOTAL IMPLEMENTED | $7,381 | - |
$3,992
108,798
kWh
|
$2,566
-
kW-mo/yr
|
$823 |