Assessment Year | 2025 |
---|---|
Principal Product | Medical Devices |
NAICS | 339113: Surgical Appliance and Supplies Manufacturing |
SIC | 3841: Surgical and Medical Instruments and Apparatus |
Sales Range | $10 M to 50M |
# of Employees | 365 |
Plant Area |
67,030
ft2
|
Annual Production | 44,656,911
units
|
Production Hrs. Annual | 6,400
hrs
|
Location (State) | New York |
Facility Energy Cost and Usage
Source | Yearly Cost | Usage | Unit Cost | |
---|---|---|---|---|
Electricity Usage | $285,886 | 5,394,233 |
kWh
|
$0.053 |
Electricity Demand | $22,242 | 2,800 |
kW-mo/yr
|
$7.94 |
Electricity Fees | $22,307 | - |
|
- |
Natural Gas | $6,833 | 925 |
MMBtu
|
$7.39 |
TOTAL YEARLY COSTS | $337,269 | |||
RECOMMENDED SAVINGS* | $249,576 |
See individual recommendations below
*Non-energy impacts included in savings.
|
||
IMPLEMENTED SAVINGS* | - |
Recommendations
# | Description | Yearly Savings* | Imp'd Cost* | Status | Electricity Usage Savings | Electricity Demand Savings | Natural Gas Savings | One-time Revenue or Avoided Cost Savings |
---|---|---|---|---|---|---|---|---|
01 | 2.4239: Eliminate or reduce compressed air usage | $2,092 | - |
$2,092
39,489
kWh
|
- | - | - | |
02 | 2.4322: Use or replace with energy efficient substitutes | $1,316 | $114 |
$1,318
24,863
kWh
|
$781
-
kW-mo/yr
|
$(-783.1)
-
MMBtu
|
- | |
03 | 4.651: Install back-up equipment | $195,000 | $86,000 | - | - | - | $195,000 | |
04 | 2.7142: Utilize higher efficiency lamps and/or ballasts | $4,092 | $3,600 |
$1,208
22,790
kWh
|
$350
-
kW-mo/yr
|
- | $2,534 | |
05 | 2.4322: Use or replace with energy efficient substitutes | $486 | $750 |
$209
3,492
kWh
|
$43
-
kW-mo/yr
|
- | $234 | |
06 | 2.6211: Conserve energy by efficient use of vending machines | $163 | $376 |
$163
3,066
kWh
|
- | - | - | |
07 | 2.2437: Recover waste heat from equipment | $4,167 | $12,180 |
$4,167
78,620
kWh
|
- | - | - | |
08 | 2.7135: Install occupancy sensors | $1,117 | $5,009 |
$1,117
21,068
kWh
|
- | - | - | |
09 | 2.9114: Use solar heat to make electricty | $41,143 | $386,971 |
$41,143
776,290
kWh
|
- | - | - | |
TOTAL RECOMMENDED | $249,576 | $495,000 |
$51,416
969,678
kWh
|
$1,174
-
kW-mo/yr
|
$(-783.1)
-
MMBtu
|
$197,768 | ||
TOTAL IMPLEMENTED | - | - | - | - | - | - | ||
Yearly Savings* | Imp'd Cost* | Electricity Usage Savings | Electricity Demand Savings | Natural Gas Savings | One-time Revenue or Avoided Cost Savings |
*Savings and costs are best estimates based on the center's site visit, associated engineering analysis, and typical capital and labor costs