Assessment Year | 2025 |
---|---|
Principal Product | Research |
NAICS | 334516: Analytical Laboratory Instrument Manufacturing |
SIC | 3674: Semiconductors and Related Devices |
Sales Range | less than $500 k |
# of Employees | 35 |
Plant Area |
75,000
ft2
|
Annual Production | 1,000
units
|
Production Hrs. Annual | 3,380
hrs
|
Location (State) | Oregon |
Facility Energy Cost and Usage
Source | Yearly Cost | Usage | Unit Cost | |
---|---|---|---|---|
Electricity Usage | $208,107 | 2,369,171 |
kWh
|
$0.088 |
Natural Gas | $32,990 | 3,875 |
MMBtu
|
$8.51 |
TOTAL YEARLY COSTS | $241,097 | |||
RECOMMENDED SAVINGS* | $9,099 |
See individual recommendations below
*Non-energy impacts included in savings.
|
||
IMPLEMENTED SAVINGS* | - |
Recommendations
# | Description | Yearly Savings* | Imp'd Cost* | Status | Electricity Usage Savings | Natural Gas Savings |
---|---|---|---|---|---|---|
01 | 2.7142: Utilize higher efficiency lamps and/or ballasts | $5,192 | $35,497 |
$5,192
59,109
kWh
|
- | |
02 | 2.7135: Install occupancy sensors | $52 | $250 |
$52
591
kWh
|
- | |
03 | 2.2511: Insulate bare equipment | $1,025 | $2,274 |
$1,025
11,665
kWh
|
- | |
04 | 2.7314: Reduce ventilation air | $2,830 | $5,000 |
$2,367
26,950
kWh
|
$463
-
MMBtu
|
|
TOTAL RECOMMENDED | $9,099 | $43,021 |
$8,636
98,315
kWh
|
$463
-
MMBtu
|
||
TOTAL IMPLEMENTED | - | - | - | - |
*Savings and costs are best estimates based on the center's site visit, associated engineering analysis, and typical capital and labor costs